Selecting the Optimum Test Tones and Test Equipment for Successful High-Speed ADC Sinewave Testing
时间:2019-12-24
大小:71.75KB
阅读数:404
查看他发布的资源
资料介绍
Abstract: An earlier application note, "Coherent Sampling vs. Window Sampling," covered the basics of coherent sampling. It showed differences between tests performed with coherent sampling and windowed sampling conditions. The following technical discussion is a follow-up note, which deals with the proper selection of test tones and instruments to successfully test and evaluate a high-speed ADC's AC performance. Maxim > App Notes > A/D and D/A Conversion/Sampling Circuits Basestations/Wireless Infrastructure
High-Speed Signal Processing
Keywords: high-speed ADCs, analog to digital converters, coherent sampling, input test tone, sinewave testing, Dec 18, 2002
clock frequency, sampling frequency, equipment, low phase noise, signal generator, logic analyzer,
synthesizer, data acquisition software, MATLAB
APPLICATION NOTE 1819
Selecting the Optimum Test Tones and Test Equipment for
Successful High-Speed ADC Sinewave Testing
Abstract: An earlier application note, "Coherent Sampling vs. Window Sampling," covered the basics of
coherent sampling. It showed differences between tests performed with coherent sampling and windowed
sampling conditions. The following technical discussion is a follow……
版权说明:本资料由用户提供并上传,仅用于学习交流;若内容存在侵权,请进行举报,或
联系我们 删除。